Introduction
High-Precision, High-Versatility Spectrophotometer
The CM-3610A Spectrophotometer is a high-precision and high-reliability bench-top instrument with vertical alignment capable to measure colours either in reflectance or transmittance and thus offering a wide range of applications such as Plastics, Textiles, Paints, Ceramics etc.
The CM-3610A utilize the patented Innovative Optical System technology including numerical gloss control (NGC) for simultaneous measurements with (SCI) and without (SCE) specular components. In addition numerical UV control (NUVC) is the unsurpassed technology for UV adjustments when measuring samples containing optical brighteners such as pulp, paper textiles and other chemicals.
These patented technologies assure highest accuracy and repeatability levels combined with superior reliability due to reduced number of moving parts and thus offer an unmatched price performance ratio.
Specifications
Model | Spectrophotometer CM-3610A | |
---|---|---|
Illumination / Viewing system | Reflectance: d:0° (diffused illumination, 0-degree viewing) Conforms to ISO 2469, JIS P8148, DIN 53145-1 and DIN 53145-2 standards. | |
Light receiving element | Silicon photodiode array (dual 40 elements) | |
Spectral separation device | Diffraction grating | |
Wavelength range, pitch | 360 nm to 740 nm | |
Wavelength pitch | 10 nm | |
Reflectance range | 0-200%, resolution 0.01% | |
Light source | Pulsed xenon lamps (× 3) | |
Measurement time | Approx. 1.5 seconds (for measurements of fluorescent colors, at 9600 bps) | |
Minimum interval between measurements | Approx. 4 seconds; when reflectance measured, approx. 5 seconds; when fluorescent color | |
Measurement/illumination area | Ø 30 mm / Ø 34 mm | |
Inter instrument agreement | Mean ΔE*ab 0.2 based on 12 BCRA Series II color tiles compared to values measured with master body | |
Repeatability | Spectral reflectance: Standard deviation within 0.1% Colorimetric values: Standard deviation within ΔE*ab 0.02 (condition; white calibration plate measured 30 times at 10-second intervals) | |
Temperature drift | Spectral reflectance: Within +/- 0.10%/°C Color difference: Within ΔE*ab 0.05/°C | |
UV adjustment | Instantaneous numerical adjustment | |
UV cut filter | 400 nm cutoff and 420 nm cutoff | |
Control method | Directly connected to a computer | |
Interface | RS-232C format | |
Power | 100-240V AC 50-60Hz 25W AC (with a dedicated AC adapter) | |
Dimensions (W x H x D) | 300 × 585 × 315 mm | |
Weight | 15.5kg | |
Operating temperature/humidity range | 13 – 33°C, relative humidity 80% or less (at 30°C) with no condensation | |
Standard accessories | White calibration plate CM-A 133, Zero calibration box CM-A119, AC adapter AC-A12, RS-232C cable IF-A12, Accessory Case CM-A117, Dust Cover CM-A118, Unit Driver CM-A108 |
Specifications are subject to change without prior notice.